53210A 53220A 53230A
Time interval A to B, B to A
Range
9
N/A -1 ns to 100,000 s (nom)
-0.5 ns to 100,000 s (min)
Time interval A or B
Range N/A 2 ns to 100,000 s (min)
Minimum width N/A 2 ns
Minimum edge repetition rate N/A 6 ns
Level & slope N/A Auto-level or user selectable
Single-period, pulse-width, rise time, fall time
Range N/A 0 s to 1000 s
Minimum width N/A 2 ns
Minimum edge repetition rate N/A 6 ns
Level & slope N/A Auto-level or user selectable
Duty
Range N/A .000001 to .999999 or 0.0001% to 99.9999%
Minumim width N/A 2 ns
Level & slope N/A Auto-level or user selectable
Phase A to B, B to A
Range
6
N/A -180.000º to 360.000º
Totalize measurements
Channels N/A Ch 1 or Ch 2
Range
9
N/A 0 to 10
15
events
Rate N/A 0 - 350 MHz
Gating N/A Continuous, timed, or external gate input
Gate accuracy is 20 ns
Level measurements
Voltage level - standard input
channels
±5.1 Vpk with 2.5 mV resolution or ±51 Vpk with 25 mV resolution
Microwave power level (micro-
wave channel option)
0 to 4 relative signal power
Measurement Characteristics (continued)
6 | Keysight | 53200A Series RF/Universal Frequency Counter/Timers - Data Sheet