70 WATTS
SINGLE/MULTI OUTPUT AC-DC
FEATURES:
CHASSIS/COVER OPEN FRAME
SAFETY SPECIFICATIONS
Underwriters Laboratories
File E137708/E140259
UL 62368-1:2014, 2
nd
Edition
CAN/CSA-C22.2 No. 62368-1-14
AAMI/ANSI ES60601-1:2005/(R) 2012
CAN/CSA-C22.2 No. 60601-1:2014
CB Reports/Certificates (including all
National and Group Deviations)
IEC 62368-1:2014, 2nd Edition
IEC 60601-1:2005/A1:2012
EN 62368-1:2014, 2nd Edition
EN 60601-1:2006/A1:2013
(2014/35/EU of February 2014)
(2011/65/EU of June 2011)
Electrical Equipment (Safety) Regulations 2016 SI No. 1101
Restriction of the Use of Certain Hazardous Substances in EEE Regulations
2012 SI No. 3032 + 2019 SI No.492
ORDERING INFORMATION
Consult factory for alternate output configurations.
Consult factory for positive, negative or floating outputs.
Please specify the following optional features when ordering:
REL-70
OUTPUT SPECIFICATIONS
Total Output Power at 50°C
(1)
Convection Cooled
(16)(18)
300LFM Forced-Air Cooled
(15)(17)(19)
(All outputs at 50% load)
Output Voltage Adjust Range
Output Overvoltage Protection
Output Overpower Protection
110-160% rated Pout, cycle on/off, auto recovery
16mS min., Full Power, 85V Input
78% Typ., Full Power, 230V, varies by model
ENVIRONMENTAL SPECIFICATIONS
Derating: See Power Rating Chart
Ambient Storage Temp. Range
2MOPP (Means of Patient Protection)
1MOPP (Means of Patient Protection)
Operational Insulation(Consult factory for 1MOPP)
Dielectric Strength
(8, 9)
5656 VDC, Primary to Secondary
2121 VDC, Primary to Ground
707 VDC, Secondary to Ground
Logic low with input power failure 10 ms
minimum prior to Output 1 dropping 1%
Remote Sense (singles only)
(10)
250mV compensation of output cable losses
Mean-Time Between Failures
100,000 Hours min., MIL-HDBK-217F, 25° C, GB
1.00 Lbs. Chassis and Cover
EMC SPECIFICATIONS (IEC 60601-1-2:2014, 4
TH
ED./IEC 61000-6-2:2005)
8KV contact / 15KV air discharge A
Radiated Electromagnetic Field
80MHz-2.7GHz, 10V/m, 80% AM A
Electrical Fast Transients/Bursts
2 KV line to earth / 1 KV line to line A
0.15 to 80MHz, 10V, 80% AM A
0% U
T
, 0.5 cycles, 0-315° 100/240V A/A
0% U
T
, 1 cycles, 0° 100/240V A/A
40% U
T
, 10/12 cycles, 0° 100/240V B/A
70% U
T
, 25/30 cycles, 0° 100/240V B/A
0% U
T
, 300 cycles, 0° 100/240V B/B
Harmonic Current Emissions
Voltage Fluctuations/Flicker
All specifications are maximum at 25°C/70W unless otherwise stated, may vary by model and are
subject to change without notice.
• Compact 2.5 x 4.5" x 1.2" Size
• 2 Year Warranty
• Universal 85-264V Input
• One to Four Outputs
• High Efficiency
• 0-70°C Operating Temperature
• IEC 60601-1 3
rd
ed. Medical Cert.
• IEC 62368-1 2
nd
Certification
• IEC 60601-1-2 4
th
ed. EMC
• Class B Emissions per EN55011/32
• RoHS Compliant
• Optional Chassis/Cover
INTEGRATED
POWER DESIGNS 300 Stewart Road ■ Wilkes-Barre, PA 18706 ■ Phone: (570) 824-4666 ■ Fax: (570) 824-4843 ■ Email: sales@ipdpower.com ■ Web: www.ipdpower.com