54
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
For email, phone or live chat, please go to
te.com/help
corcom.com
Multi-purpose Medical Filter for Power Line Noise Protection
MV Series
MV Series
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz: .07 mA
@250 VAC 50 Hz: .13 mA
Hipot rating (one minute):
Line to Ground: 2250 VDC
Line to Line: 1450 VDC
Rated Voltage (max): 250 VAC
Operating Frequency: 50/60 Hz
Rated Current: 3 to 20A
Operating Ambient Temperature Range
(at rated current I
r
): -10°C to +40°C
In an ambient temperature (T
a
) higher than +40°C
the maximum operating current (I
o
) is calculated as
fo
llows: I
o
= I
r
(85-Ta)/45
• Multi-purpose medical filter
• Improved Line to Ground performance
• A good solution to emission or immunity problems
• Meets leakage current requirements of UL2601 for
health care equipment
Ordering Information
6 MV 1
Input / Output Style
1 – .250 [6.3] spade terminals
M
V Series
Current Rating
3, 6, 10 or 20A
Specifications
UL Recognized
CSA Certified
VDE Approved
Available Part Numbers
3MV1 6MV1
10MV1 20MV1
Electrical Schematic