43
RFI Power Line Filters
High Performance Power Line Filter for Medical Applications
1
Dimensions are in inches and millimeters unless otherwise specified. Values in italics
are metric equivalents. Dimensions are shown for reference purposes only.
Specifications subject to change.
Corcom Product Guide
Catalog: 1654001
Issue Date: 06.2011
For email, phone or live chat, please go to
te.com/help
corcom.com
HZ Series
HZ Series
Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz: 2 A
@250 VAC 50 Hz: 5 A
Hipot rating (one minute):
Line to Ground: 2250 VDC
Line to Line: 1450 VDC
Rated Voltage (max): 250 VAC
Operating Frequency: 50/60 Hz
Rated Current: 3 to 30A
Operating Ambient Temperature Range
(at rated current I
r
): -10°C to +40°C
In an ambient temperature (T
a
) higher than +40°C
the maximum operating current (I
o
) is calculated a
s
fo
llows: I
o
= I
r
(85-Ta)/45
• Designed to provide good attenuation to RFI noise
in the frequency range from 10kHz to 30MHz
• Size and cost-effective
• Low leakage current
• New versions up to 30A
Ordering Information
3 E HZ 1
Input / Output Style
1 – .250 [6.3] spade terminals
6 – Threaded bolt
H
Z Series
Leakage current designation
E – Low leakage (<0.5 mA)
Current Rating
3, 4, 6, 10, 15, 20 or 30A
Specifications
UL Recognized
CSA Certified
VDE Approved
Available Part Numbers
Electrical Schematic
LOAD
3EHZ1 4EHZ1
6EHZ1 10EHZ1
15EHZ1 20EHZ1
30EHZ6
Case Styles
A
B
C
D
E
3EHZ1